Ductless fume hoods, Chemcap™ Clearview™

Supplier: Bigneat
BIGNBC1004/AM/HPEA 4190 EUR
BIGNBC1004/AM/HP BIGNBC4002/HPLUS BIGNBC6003/HPLUS BIGNBC1204/AM/HP BIGNBC8004/HPLUS BIGNBC8004/F/HP 135-1987 BIGNBC4002/HPL/HP 135-1986 135-0650 135-1981 135-1980 135-1983 135-0652 135-0651 135-1982 135-1985 135-0654 135-0653 135-1984 BIGNBC6003/AM BIGNBC1204/F/HP BIGNBC1004/HPL/HP BIGNBC8004/HPL/HP 135-1976 135-0644 135-1975 135-1978 135-0646 135-1977 135-0649 135-1979 BIGNBC6003/HPL/HP BIGNBC1206-4 135-1970 135-1972 135-1971 135-2587 135-1974 135-0643 135-1973 BIGNBC1004/HPLUS 135-1965 135-1964 135-1967 BIGNBC4002/F/HP 135-1966 135-1969 BIGNBC1204/HPLUS BIGNBC0603-4 135-1968 135-1961 135-1960 135-1963 135-1962 BIGNBC1506-4 BIGNBC1506/HP BIGNBC4002/AM/HP BIGNBC8004/AM/HP BIGNBC6003/AM/HP 135-1959
Ductless fume hoods, Chemcap™ Clearview™
Fume Hoods and Enclosures Ductless Fume Hoods
Chemcap™ Clearview™ ductless fume hoods include Chemcap™ carbon filtration. A range of six standard cabinet sizes with keypad control panel, touch button controls and service due indicator. Includes 18 W fluorescent lamp and PVC base tray which offers excellent chemical resistance and contains spillages.
Cabinets are manufactured from polyester coated steel; transparent panels are acrylic and resistant to corrosion.

  • Fully portable
  • Clearview all-round vision
  • BC 4002 has a full width envelope access
  • BC 6003 to BC 1204 have teardrop front working access
  • BC 1206 and BC 1506 cabinets have double fan, double filter stack and have trapezium front working access

Operator protection: Chemcap™ fume cabinets protect the operator from fumes, aerosols and/or particulates whilst undertaking laboratory procedures.

Chemcap™ carbon/HEPA filtration options: Filters are included - most frequently used MP (multipurpose) and OS (where organic fumes predominate) are listed below, with or without HEPA filtration. Alternative filter types H+ (acids), F (formaldehyde), AM (ammonia) are available on request.

Power supply: 230 V, AC, 50 Hz, 5 Amp, 1 Ø (on request 110 V, AC, 60 Hz, 8 Amp, 1 Ø).

Certifikace: Chemcap™ Clearview™ conforms to the requirements of all major standards for containment and filtration: US Standard ASHRAE 110: 1995 12-203 (for containment)
British Standard BS 7989:2001 (for filter efficiency and retention capacity). Optional rear baffle is required, please contact Avantor for details.
French Standard AFNOR NFX 15-211:2009 - Class 1 (for filter retention capacity)

Informace pro objednávání: Please contact Avantor for details on alternative filter types and other options, including mobile stands and service options.

Visit by qualified engineer to deliver, assemble, commission and air flow test powder weighing, ductless fume or PCR cabinets. Advice to user(s) of best practice.
Order Now

Learn more

About VWR

Avantor is a vertically integrated, global supplier of discovery-to-delivery solutions for...

Learn more About VWR